<Error>
<ErrorNumber>47400</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: Picker was unable to measure stack heights in cassette &lt;Cassette&gt; because the height differential between each side was greater than &lt;Maximum differential&gt; mm
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The picker measured the media stacks but the difference in height between the two stacks
to be greater than the allowable maximum.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Remove or add media from the cassette specified so that the stack height differential is
below the maximum specified. Check that the media is correctly specified in the cassette
(one stack vs two stacks).
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Operator misloaded the cassette
</Cause>
<Parameter>
Cassette -
This value indicates which cassette was being measured
</Parameter>
<Parameter>
Maximum differential -
This value indicates the maximum stack height differential for measuring in mm
</Parameter>
<LastModified>
28-Apr-2008
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:51 2008 PST
</Updated>
<Links>
</Links>
</Error>